<ErrorNumber>27006</ErrorNumber>
<ErrorFamily>Debris Removal</ErrorFamily>
<ErrorMessage>
DEB: The flow rate is below the minimum required
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>UDRC</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
If you can hear the cabinet's blower running and the filter is new (or clean) this may indicate a blocked vacuum sensor port on the debris nozzle. Please clear the vacuum sensor port using this procedure http://techplanet.creo.com/output/heads/thermal/content/docs/73-3667A-C-PR01.pdf . To prevent repeated blockage of the nozzle please check to ensure that the current version of the debris nozzle is installed and, if applicable, an airjet plate is installed.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
</text></DescriptionText>
</OperatorAction>
<Cause>
Cause 1:  Voltage too low at the cabinet. Cabinet has been moved to a lower voltage location or long extension cord has been added. Minimum operating voltage is 190V at the cabinet.
Solution: Increase voltage
</Cause>
<Cause>
Cause 2:  Sudden plugging of filter.
Solution: Replace particle filter
</Cause>
<Cause>
Cause 3:  Mechanical damage to blower
Solution: Replace blower
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:34:13 2008 PST
</Updated>
<Links>
</Links>
</Error>